Gene/Protein
Disease
Symptom
Drug
Enzyme
Compound
Pivot Concepts:
Gene/Protein
Disease
Symptom
Drug
Enzyme
Compound
Target Concepts:
Gene/Protein
Disease
Symptom
Drug
Enzyme
Compound
Query: UNIPROT:Q9BZ95 (
Whistle
)
42
document(s) hit in 31,850,051 MEDLINE articles (0.00 seconds)
A redesigned Peak Flow
Whistle
is described. It was evaluated and compared to the
Wright
Peak Flow Meter (WPFM) and the Vitalograph Pulmonary Monitor (VPM) using a laboratory flow system. These devices were evaluated for accuracy at steady flows, resistance characteristics, measurement of drift and dynamic accuracy. All devices tested correlated to a significant degree for accuracy at steady flows. The resistance of the WPFM and VPM increased with higher flows, whereas the Cardboard Peak Flow
Whistle
(CPFW) decreased. After pulsing the CPFW did not show any drift in calibration. The dynamic accuracy of the CPFW was excellent. Thus the accuracy of the CPFW and its lower cost makes it a practical device for measuring peak flows.
...
PMID:Report of a newly redesigned Peak Flow Whistle. 670 12
A peak flow (PF) measuring device, the Peak Flow
Whistle
(PFW), has recently been designed. The principle of a threshold activated reed is utilized to register PF, which differs significantly from the spring distortion mechanism utilized in currently marketed devices. Prototypes of the PFW were tested in the laboratory and in clinical settings as part of our preliminary studies. When compared with a pneumotachograph in the laboratory, the values correlated significantly (p less than .0001). In field studies both in asthmatic and non-asthmatic children, the PFW values correlated significantly (p less than .0001) with those of the
Wright
Peak Flow Meter and performance was found to be satisfactory.
...
PMID:Peak flow whistle: preliminary report. 725 46