Gene/Protein Disease Symptom Drug Enzyme Compound
Pivot Concepts:   Target Concepts:
Query: UMLS:C0033377 (prolapse)
11,717 document(s) hit in 31,850,051 MEDLINE articles (0.00 seconds)

While modern implant tools have contributed greatly to the success of cardiac resynchronization therapy, technical challenges remain. A common problem is the inability to advance left ventricular pacing leads into branch veins that are tortuous or arise at steep angles. In these cases, advancement of the lead causes it to buckle and prolapse into the coronary sinus or great cardiac vein. Lead prolapsed can be avoided by employing a balloon to temporarily obstruct the coronary sinus or great cardiac vein just upstream from the branch vein. The balloon redirects the force of advancement laterally into the branch vein, facilitating delivery.
...
PMID:Balloon-facilitated delivery of a left ventricular pacing lead. 2127 22