Gene/Protein
Disease
Symptom
Drug
Enzyme
Compound
Pivot Concepts:
Gene/Protein
Disease
Symptom
Drug
Enzyme
Compound
Target Concepts:
Gene/Protein
Disease
Symptom
Drug
Enzyme
Compound
Query: UMLS:C0021051 (
immunodeficiency
)
71,517
document(s) hit in 31,850,051 MEDLINE articles (0.00 seconds)
We compared the
Abbott
enzyme immunoassay for human
immunodeficiency
virus (HIV) antigen with the reverse transcriptase assay (RTA) as a means of monitoring HIV infection during an antiviral trial. The
Abbott
enzyme immunoassay detected HIV earlier than RTA whether or not the patients were antigenemic and appears to be superior to RTA for detecting HIV in cultures used for monitoring clinical trials.
...
PMID:Comparison of antigen immunoassay and reverse transcriptase assay for monitoring human immunodeficiency virus infection in an antiviral trial. 246 May
The detection and recruitment of HIV antigen-positive asymptomatic individuals for clinical trials is important. Two commercial enzyme-linked immunosorbent assays (ELISA) for the detection and quantitation of human
immunodeficiency
virus (HIV) antigens were evaluated for sensitivity by testing serum samples from 155 asymptomatic HIV Western blot positive individuals. The
Abbott
HIV antigen ELISA detected HIV antigen in the serum of 17 (11.0%) of 155 patients compared with 18 (11.6%) of 155 by the Coulter HIV antigen ELISA. In serial twofold dilution experiments, there was no significant difference in sensitivity between these two assays in the detection of HIV serum antigen. However, both assays are limited in their ability to detect HIV antigen in most asymptomatic HIV-infected patients. This low detection rate should be taken into account in the design of clinical trials involving asymptomatic infected patients.
...
PMID:Comparison of two serum HIV antigen assays for selection of asymptomatic antigenemic individuals into clinical trials. 250 16
Three enzyme immunoassay (EIA) methods for the detection of human
immunodeficiency
virus (HIV-1) were evaluated. Serum or plasma samples from 22 individuals seropositive for HIV-1 antibodies were tested with the
Abbott
, Coulter, and DuPont kits for presence of HIV-1 p24 antigen. Another 12 samples were tested with two kits only. Discordant results were obtained with 9 of 34 (26%) HIV-1-antibody-positive patient samples tested. Most of these discrepancies were found in samples containing less than 30 pg/ml of HIV-1 p24 core antigen. A sampling of sera from normal blood donors and patients with infectious or autoimmune diseases revealed a low level of false positive reactions, especially with sera containing antinuclear antibodies or rheumatoid factor. Noteworthy is the frequency of false positive reactions seen with the DuPont EIA for HIV-1 p24 antigen. 18/111 sera (16.2%) containing auto-antibodies tested positively with the DuPont HIV-1 p24 antigen EIA. The nonspecific nature of the test reactivity for 9/10 of these samples was confirmed using an HIV-1 p24 antigen inhibition assay. These findings are discussed in light of the need for HIV-1 antigen detection in the clinical laboratory and of other methods for HIV-1 detection: the polymerase chain reaction and measurements of reverse transcriptase activity.
...
PMID:Evaluation of three enzyme immunoassays for HIV-1 antigen detection. 251 47
A flow cytometric immunofluorescence assay (FIFA) was developed to detect antibodies to human
immunodeficiency
virus (HIV) using live cell indirect immunofluorescence and analysis by flow cytometry. A panel of 107 sera, previously tested for anti-HIV antibody with the
Abbott
Enzyme-Linked Immunosorbent Assay test and Western blot (WB), was rescreened by FIFA. Antibodies were tested on HIV-infected and uninfected H9 cells in the FIFA. Although ELISA results indicated seven false positive results by comparison with the WB, 46 of 46 FIFA positive results tested WB positive and 61 of 61 FIFA negative results were WB negative. The results of FIFA showed 100% sensitivity and 100% specificity compared with WB. This rapid, quantitative, relatively easy assay makes FIFA an alternate confirmatory test for the presence of HIV antibodies.
...
PMID:Flow cytometric indirect immunofluorescence assay with high sensitivity and specificity for detection of antibodies to human immunodeficiency virus (HIV). 264 8
The high prevalence of human
immunodeficiency
virus (HIV-1) infection in populations at risk in Miami prompted a seroepidemiologic study of both HIV-1 and the human T-cell leukemia virus type I (HTLV-I), a closely related virus, in our patients receiving chronic hemodialysis. One hundred twenty-nine patients undergoing hemodialysis in 1986 and 1987 were tested for antibody against both viral antigens by EIA (
Abbott
Laboratories,
Abbott
Park, IL). Seroreactive samples for HIV-1 and/or HTLV-I were confirmed by Western blot and, for HTLV-I, by viral cultures. Thirty patients (23.2%) were positive for retroviral infection (22 for HIV-1 alone, four for HTLV-I alone, and four for both HIV-1 and HTLV-I). The most important risk factor was intravenous drug use, followed by blood transfusion. Patients with HIV-1 had lower T4-T8 ratios and higher mortality than those with HTLV-I infection alone. It was concluded that HTLV-I, as well as HIV-1, infection is endemic in chronic dialysis centers in Miami. The clinical consequences of HTLV-I infection in relatively immunocompromised patients with chronic uremia who are undergoing chronic hemodialysis remains to be established.
...
PMID:Human immunodeficiency virus and human T-cell leukemia virus type I in patients undergoing maintenance hemodialysis in Miami. 278 17
Six commercial enzyme immunoassays (EIA) were evaluated in 6488 serum samples, with immunoblot analysis as the confirmatory test for antibodies against human
immunodeficiency
virus (HIV). The
Abbott
and Wellcome tests identified all 163 immunoblot-positive samples correctly, whereas the other tests did not detect 1-3 samples. In AIDS patients (predominantly with antibodies to gp41env) Organon's EIA was less sensitive (p less than 0.05) and Wellcome's more sensitive (p less than 0.05) than the immunoblot assay. In symptom-free anti-HIV-positive subjects (antibodies to almost all viral antigens and high titres of anti-p24gag) all the EIA were significantly (p less than 0.05) less sensitive than the immunoblot assay. The frequencies of false-positive reactions in a "tricky" panel of samples from patients with autoimmune and acute viral diseases and in a blood-donor panel were
Abbott
9.5%, 0.42%: Organon 1.7%, 0%; Litton 1.0%, 0.4%; Behring 2.7%, 0.06%; Wellcome 0%, 0%; and Pasteur 0%, 0.02%. The results of a seventh EIA (Dupont) were excluded from the study at the company's request. All six EIA evaluated are suitable tests for anti-HIV screening in samples from patients and blood donors.
...
PMID:Evaluation of six enzyme immunoassays for antibody against human immunodeficiency virus. 287 39
One hundred and thirty-five sera were tested for the presence of antibody to human
immunodeficiency
virus (HIV) by enzyme immunoassay (EIA) with the
Abbott
Diagnostics HTLV-III test before and after heating at 56 degrees C for 10 min. Only one serum specimen was repeatedly reactive after heating when compared with the unheated portion (1/135). The specimen was a low level positive (ratio of test over cutoff = 1.6). Fifty-eight different sera, selected to yield a high percentage of positive results, had both their heated and unheated Western blot (WB) results agree completely with repeated EIA testing. An additional 4,244 sera heated at 56 degrees C for 10 min were tested. Of the 330 repeatable EIA positives, only 38 were not confirmed by WB. HIV infectivity in sera was reduced from 10(3.5) TCID50 to 10(1) TCID50 by the same heat treatment. We conclude that heating sera at 56 degrees C for 10 min significantly reduces HIV infectivity and does not significantly affect the results of HIV antibody testing.
...
PMID:Effect of 10 minutes heat treatment on HIV antibody testing from alternate testing sites. 305 10
The authors have carried out, on 150 sera of patients seropositive for the human
immunodeficiency
virus type I (HIV I) and 11 cerebrospinal fluid of which 5 were patient infected by the HIV I, a comparative study of two commercial tests for the detection of HIV I antigen (Diagnostic Pasteur and
Abbott
laboratories). A much greater sensitivity was obtained with the specificity being practically identical for the sera with the two tests (100% with
Abbott
laboratories test, 96.11% with the diagnostic Pasteur test). 4 sera appeared "false negatives" with the
Abbott
Laboratories test; their optical density was situated between 80 and 100 p. cent of the cut-off level value, whereas that of the "real" negatives was situated between 30 and 60 p. cent of the cut-off level value. 10 of the 11 cerebrospinal fluids appeared false positive with the Diagnostic Pasteur. This seems to be connected with an insufficiency of saturation of protein receptors in the wells. The Diagnostic Pasteur test is not adapted for the detection of HIV I antigen in the body fluids with a weak protein concentration. Contrary to the results obtained with the Encavor test (
Abbott
laboratories) the analysis in western-blot does not show an inverse prevalence of anti p24 GAG antibodies with regard to antigen HIV I in seropositive patients. On the other hand, the statistical analysis of the positive HIV I sera which are at the same time antigen HIV I positive and antibodies HIV I positive suggests an earlier disappearance of anti p17 GAG antibodies than of anti p24 GAG antibodies.
...
PMID:[Comparative study of 2 commercial tests for the detection of HIV-1 antigens in blood and cerebrospinal fluid using immunoenzymology]. 306 38
The vitreous humor from 11 patients with acquired immunodeficiency syndrome was obtained at postmortem examination and tested for human
immunodeficiency
virus antigen and antibody by using the
Abbott
enzyme-linked immunosorbent assay procedures. Five patients had detectable antigen, supporting the recent observation that the virus may directly infect the retina.
...
PMID:Detection of human immunodeficiency virus antigen in vitreous humor. 323 68
During 1985 and 1986, the Centers for Disease Control conducted eight surveys to evaluate laboratory performance in testing for antibody to human
immunodeficiency
virus (HIV). (Other names for this virus include "human T-lymphotropic virus type III" and "lymphadenopathy associated virus.") The first survey was conducted with ten samples and 50 laboratories; the remaining surveys used six samples, but enrollment increased to as many as 475 laboratories. The purpose of these surveys was to measure test performance by the laboratories under the actual conditions of use. The surveys contained duplicate samples to permit measurement of within- and between-survey reproducibility. One survey included positive and negative reference materials as evaluation samples. Results of the Western blot (WB) test were not always positive on samples that were positive by enzyme immunoassay (EIA), and sometimes they were positive on samples that had negative results by EIA. The percentage of positive results reported by laboratories using Electronucleonics tests was lower than that reported by users of
Abbott
and of Litton tests. positive results on the negative reference material were reported by 13.7% of the
Abbott
EIA test users. The tests do not seem to be calibrated against an equivalent standard. Within-survey reproducibility was usually above 95% for EIA and WB methods and for the major manufacturers of EIA tests. Between-survey reproducibility was usually above 90%. Reproducibility was below 75% for some combinations of samples, method, and manufacturer. The test performance observed in these surveys may be lower than is actually achieved on patient samples because the samples were selected to measure technical competence and contain a higher frequency of samples with low reactivity than would be encountered with clinical samples. Stratification and weighting of the results in these surveys to estimate performance parameters for a blood bank population indicate that the tests are performing well in this application. If single EIA tests at the performance levels achieved in these surveys were used (no repeat and no confirmation), about 1 sample with positive results in every 50,000 samples tested would be missed and about 2.5% would be false positive results. Because tests are duplicated and confirmed in blood banks, better performance would be expected. Testing for antibody to HIV virus continues to undergo rapid change because of the introduction of new tests, changes in technical aspects of existing tests, and application of testing to changing populations. Such dynamic factors necessitate an ongoing, comprehensive monitoring of test performance.
...
PMID:Summary of the Centers for Disease Control human immunodeficiency virus (HIV) performance evaluation surveys for 1985 and 1986. 327 37
<< Previous
1
2
3
4
5
6
7
8
9
10
Next >>