Ammonia removal from a recirculating dialysate stream is a major challenge in developing a truly portable, regenerable hemodialysis system. Three zeolites, type F, type W, and clinoptilolite, were found to have good
ammonia ion
exchange capacity with linear equilibrium ion exchange coefficients of 0.908, 0.488, and 0.075 L/g, respectively. The linear equilibrium ion exchange coefficient relates dialysate ammonia concentration (mumol/L) to the amount of ammonia absorbed by zeolite (mumol/g) at equilibrium. Ammonia uptake by zeolite powders was fast, with equilibrium reached within 15 sec. Zeolite
ammonia ion
exchange and regeneration through multiple cycles was studied using an ion exchange column containing clinoptilolite pellets. Zeolite ion exchange capability was regenerated by
flushing
the column with 2 mol/L sodium chloride after an ion exchange run. The column maintained
ammonia ion
exchange capacity through six ion exchange/regeneration cycles, demonstrating multiple dialysis use possibilities. Atomic absorption spectroscopy of the column effluent showed no detectible (< 1 part per million) Si or Al leached from the zeolite.
...
PMID:Zeolitic ammonium ion exchange for portable hemodialysis dialysate regeneration. 858 74